MedVue Firmware Update

To download files: Click on the link above.

Revision 1.01.08 - Effective May 8, 2014

Rev 1.01.08 - Fixed issue with Height/BMI locking with weight when lock key pressed or auto-lock occurred.

Rev 1.01.07 - Update to fix production issue with MV2. Update for new MVWIFI option card hardware.

Rev 1.01.06 - Update MedVue application software to fix an issue with the Ethernet/WiFi option card. Adjust display contrast to help with the new white backlight. Improvements to help extend battery life.

Rev 1.01.05 - Update MedVue application software adding OIML Yes/No prompt to prepare for MV2 production.

Rev 1.01.04 - Update MedVue application software to fix a bug when communicating with Welch Allyn devices.

Rev 1.01.03 - Update MedVue application software to fix a bug when option card installed that would affect the display.

Rev 1.01.02 - Update MedVue application software to fix a bug in detecting the SDCard.

Rev 1.00.01 - Update MedVue application software to fix a bug in serial/TCP/USB communication.

Rev 1.00.00 - Initial software release.

Software Update Instructions:

In order to update the MedVue indicator you will need a USB-A to USB-B cable, a Phillips head screwdriver, and a paper clip.

Software Update Procedure:

• Download the current MedVue software.
• Unzip the downloaded file by right clicking and select “Extract All”.
• Using the Phillips head screwdriver, remove the left end cap of the indicator.
• Using the paper clip, press the calibration switch and power on the indicator.
• The indicator should now display “boot rev 1.0” at the bottom of the display.
• Connect the indicator to your PC using the USB-A to USB-B cable.
• You should see a new removable storage device appear in My Computer.
• Find the unzipped file from step 2 above, it should be named “MedVue.afx.S19”.
• Drag and drop the file to the new removable storage drive named “MedVue”.
• The indicator display should show “Flashing S19” while the update is progressing.
• Once done, the indicator should reset and the new revision number should appear during startup.
• If you have difficulty, try starting over from the beginning of these steps.